Senseca Germany GmbH

DX 115 - Digital Pt100 immersion probe, large handle (e.g. for PRO D0x)

Producer

DX 115 is a high-precision, digital, wire-wound Pt100 temperature probe for the PRO D01 / PRO D05 hand-held measuring devices.

The probe is calibrated at the factory and is ready for immediate use. It can optionally be supplied with a DAkkS/ACCREDIA calibration certificate.

The calibration data is stored in the probe, allowing interchangeability without recalibration.

The user can adjust the probe in two points (offset and slope) or by setting R0 (sensor resistance at 0 °C) and the Callendar-Van Dusen parameters.
Supplied with 1, 2 or 5 m cable with 4-pin M12 connector.

Protection class:
IP67
Power consumption:
<1 mA typ.
Sensor:
Pt 100 (wire wound)
Resolution:
0.01 °C
Accuracy:
±0.05 °C (t = 0 °C) ±0.1 °C (0 °C ≤ t ≤ 100 °C) ±0.2 °C (-50 °C ≤ t < 0 °C, 100 < t ≤ 250 °C) ±0.3 °C (t = remaining range)
Handle material:
Polyamide (PA6-GF30), gland Santoprene (both up to +120 °C)
Handle length:
130 mm (incl. grommet)
Response time (T63):
3 s
 
 
Enquire for more information
please weait… Your data is being loaded...

Article no. Order Code Connection Probe length Cable and length Measuring range Accuracy Version Resolution Model Calibration Diameter probe tube Sensor Weight Probe tube material Scope of delivery Measuring element Reaction time T90 Handle length minimal quantity Product comparison
486229 DX115-00-300-L02 DX115-00-300-L02 M12A connector, 4-pin, electronic in plug 300 mm 2 m PVC cable, -20 ... +105 °C -196 ... +500 °C ±0.05 K @ 0°C, 0.1 K in the range of 0...100°C, resolution 0.01°C Interchangeable without loss of accuracy 0.01 °C Immersion probe Factory adjustment, exchangeable Ø3 mm Pt100 4-wire, mineral insulated ca. 110 g incl. 2 m cable V4A Device in black protective bag 460x85x55 mm with velcro closure, Quickstart Pt100, wire wound, high precision 3 s 130 mm incl. grommet 1 Delivery time
487740 DX115-00-230-L02 DX 115-00-230-L02 M12A connector, 4-pin, electronic in plug 230 mm 2 m PVC cable, -20 ... +105 °C -196 ... +500 °C ±0.05 K @ 0°C, 0.1 K in the range of 0...100°C, resolution 0.01°C Interchangeable without loss of accuracy 0.01 °C Immersion probe Factory adjustment, exchangeable Ø3 mm Pt100 4-wire, mineral insulated ca. 110 g incl. 2 m cable V4A Device in black protective bag 460x85x55 mm with velcro closure, Quickstart Pt100, wire wound, high precision 3 s 130 mm incl. grommet 1 Delivery time
487742 DX115-I3-150-L02 DX115-I3-150-L02 M12A connector, 4-pin, electronic in plug 150 mm 2 m PVC cable, -20 ... +105 °C -196 ... +250 °C ±0.05 K @ 0°C, 0.1 K in the range of 0...100°C, resolution 0.01°C Interchangeable without loss of accuracy 0.01 °C Insertion probe Factory adjustment, exchangeable Ø3 mm Pt100 4-wire ca. 110 g incl. 2 m cable V4A Device in carton box, incl. Quickstart Pt100, wire wound, high precision 3 s 130 mm incl. grommet 1 Delivery time