Senseca Germany GmbH

PRO 13x.2 - Precision second thermometer, graphic display

Producer

INTRODUCTION
PRO 131.2 and PRO 135.2 are dual channel high class professional handheld meters for thermocouple probes with a rich set of features, high grade robustness and operating comfort for safe and reliable use.
PRO 135 also has data logging capability.

FEATURES
Display
The multilingual large dot matrix/clear text LCD has ergonomic wide-angle visibility from daylight to darkness, thanks to the backlight.
It displays either large scale values, statistical data or the chart of a variable measurement history.
The HOLD feature allows freezing the measurements on display, while the REL feature allows showing the measurement against the measured value.
°C, °F and K units of measurement are available.
If two probes are connected, the instrument calculates and shows the difference between the measurement of the two channels.
On demand, the display of the internal device temperature (reference junction) can also be switched on.
Data Logging (only PRO 135.2)
Large storage capacity: up to 1 million data, file system based.
The logged data are store in CVS files that can be easily viewed connecting the instrument to a PC via USB:
the instrument is seen by the PC as a mass storage device, the data can be read oud and evaluated without software necessarily needed.
Automatic log with configurable interval.
The instruments integrate a Real Time Clock: date and time of each logged sample are stored.
Alarm
Configurable alarm thresholds and optionally hysteresis can be set. LCD indication and buzzer activation when thresholds are exceeded.

CONFIGURATION & MEASUREMENT
Probes
Any K, J, T, N, R, S, B and E standard thermocouple probe can be connected.
Wide range of probes available.
Configurable surface correction factors can be set for use in combination with Senseca contact probes, for the best measuring accuracy.
Connection to PC
Via the USB C port, for viewing or downloading the files stored in the instrument internal memory (only PRO 135.2) or connecting to the application software ProXware.
Statistics
Detection of MIN, AVG (average) and MAX. The user can clear the statistical info to start a new statistical calculation.

Protection class:
IP 67 (except probe connection), IK 04
Storage capacity:
only PRO 135.2: Up to 1 million data sets, file system based
Each data set includes date/timestamp and measurement.
Data are stored in CVS files.
Logging type:
only PRO 135.2: Automatic with manual start/stop
Accuracy:
±(0.2 °C + 0.2% of measured value) ± 1 digit (J, K, N, T, E)
±(0.5 °C + 0.4% of measured value) ± 1 digit (S, B, R)
Measurement rate:
4 meas/s
Long-term drift:
±0.1 °C/year
Logging interval:
only PRO 135.2: 1, 2, 5, 10, 15, 30 s / 1, 2, 5, 10, 15, 30 min / 1 hour
Clock:
User-adjustable date and RTC function (max. drift 1 min/month @ 25 °C)
User interface:
Multilingual (de, en, it, fr, es)
PC connection:
USB C (cable included), Mass Storage Device
Power supply:
4 x AA battery (included in the scope of delivery)
External 5 V DC via USB C (power adapter or PC USB port)
Power consumption:
20 mA typ.
Weight:
320 g incl. battery
Battery autonomy:
> 150 h typ. continuous operation
Auto power off:
Yes, user configurable
Automatically disabled if external power is connected
Housing material:
Shock and impact proof, ABS, TPE (side protection), Polyster (front panel),
Measuring range:
K: -220.0…+1372.0 °C / -364.0…+2501.6 °F
J: -200.0…+1100.0 °C / -328.0…+2012.0 °F
N: -200.0…+1300.0 °C / -328.0…+2372.0 °F
S: -50.0…+1750.0 °C / -58.0…+3182.0 °F
T: -200.0…+400.0 °C / -328.0…+752.0 °F
E: -60.0…+850.0 °C / -76.0…+1562.0
Operating conditions:
-5 ... +50 °C; 0 ... 85 % RH (non-condensing)
Storage temperature:
-25…65 °C (without batteries)
Display:
140 x 160 dot-matrix LCD with backlight / visible area 42 x 50 mm, display of large-digit single values, statistical info (min/avg/max), chart view
Dimensions:
170 x 78 x 38 mm
Standard functions:
Acostic/optic alarm, automatic switch-off freely configurable by the user (automatically deactivated if external power supply is connected)
Accuracy typ.:
±(0.2 °C + 0.2% of measured value) ± 1 digit (J, K, N, T, E)
±(1 °C + 0.4% of measured value) ± 1 digit (S, B, R)
Battery life:
> 150 h typ. Continuous operation
Current consumption:
20 mA typ
Data logger with read out via USB:
only PRO 135.2
Extras:
Foldable stand and magnet for flexible operation
Measuring:
2 Miniature female flat-pin TC connectors, internal temperature
Optional special functions:
Favorites menu, multilingual
Resolution:
0.1 °C / 0.1 °F o 1 °C / 1 °F depending on TC type (see meas. ranges)
Cold junction accuracy:
±0.3 °C
Channels:
2
Miniature female flat-pin TC connectors
Protection degree:
IP 67 (except probe connection)
IK 04
Display units:
°C, °F, K
Max cable length:
<30 m
Ranges:
K: -220.0…+1372.0 °C / -364.0…+2501.6 °F
J: -200.0…+1100.0 °C / -328.0…+2012.0 °F
N: -200.0…+1300.0 °C / -328.0…+2372.0 °F
S: -50…+1750 °C / -58…+3182 °F
T: -200.0…+400.0 °C / -328.0…+752.0 °F
E: -60.0…+850.0 °C / -76.0…+1562.0 °F
B: +30
Supporting thermocouple types:
K, J, T, N, R, S, B, E
 
 
Enquire for more information
please weait… Your data is being loaded...

Article no. Version Resolution Measuring range Probe connections Set-Option Display Current supply Alarm function Brand Scope of delivery Measuring input Output Data logger Protection type minimal quantity Quantity
PCE
486118 PRO135.2 With data logger 0.1 °C / 0.1 °F -220 ... +1768 °C (depends on the used thermocouple) 2 sockets for 2-pin standard flat plugs Device in carton box, testprotocol, Quickstart, without temperature probe Graphic display with backlight 4 batteries (Mignon AA/LR6) or via USB socket CH1, CH2, DIF, sound & light SENSECA incl batteries, short manual 2x thermocouples type J, K, N, S, T, B, E, R USB jack type C Up to 1 million data sets, file system based IP67 (except sensor input) 1 Delivery time
486119 PRO131.2 Standard 0.1 °C / 0.1 °F -220 ... +1768 °C (depends on the used thermocouple) 2 sockets for 2-pin standard flat plugs Device in carton box, testprotocol, Quickstart, without temperature probe Graphic display with backlight 4 batteries (Mignon AA/LR6) or via USB socket CH1, CH2, DIF, sound & light SENSECA incl batteries, short manual 2x thermocouples type J, K, N, S, T, B, E, R USB jack type C   IP67 (except sensor input) 1 Delivery time